RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3214035379 - Archived data from Fri, 05. February 2021 11:06:34 CET


This is all the information about APD 3214035379. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3214035379
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: G07
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 385.1 V
Dark current: 25.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 141
Position in Box: 43
EP1 batch: 164
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 383.7630178 V     T = -25 °C: 348.1372378 V
Voltage for Gain 150: T = +20 °C: 391.6409953 V     T = -25 °C: 355.5733587 V
Voltage for Gain 200: T = +20 °C: 396.0118947 V     T = -25 °C: 359.7502087 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.658257558 V-1     T = -25 °C: 4.840784588 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.212227881 V-1     T = -25 °C: 9.75410859 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.37497853 V-1     T = -25 °C: 15.46050348 V-1
Break-through voltage: T = +20 °C: 410.8227227 V     T = -25 °C: 374.4411672 V
   
Notes:


Version history

Time Author Change comment
15. Nov 2022 13:09:10 CET oafedulidis Assigned to capsule by script
30. Aug 2021 14:07:11 CEST oafedulidis Updated box and position via API.
05. Feb 2021 11:06:34 CET jreher Updated box and position via API.
11. Aug 2020 17:55:02 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:51:26 CEST jreher Updated APD location via API.
12. Dec 2019 14:41:08 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:48:42 CET jreher Updated APD location via API.
27. Nov 2019 19:53:46 CET jreher Updated APD location via API.
25. Nov 2019 17:22:06 CET jreher Updated box and position via API.
20. Sep 2019 18:49:46 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:55:07 CEST jreher Updated APD location via API.
03. Jan 2019 00:15:04 CET Tobias Corrected dark current unit.
14. Aug 2018 16:04:24 CEST Tobias Imported from Hamamatsu datasheet.