Show Hamamatsu Avalanche Photo Diode 2702030075
					
					This is all the information about APD 2702030075. If it is wrong, edit the data.
					
						
							| Subdetector specification: | 
							  | 
						
						
							| Serial: | 
							2702030075 | 
						
						
							| Type: | 
							Hamamatsu Avalanche Photo Diode | 
						
						
							| Detector: | 
							assigned to 2705030345/2702030075 | 
						
						
							| Unit: | 
							#1700 (barcode 1309018192) | 
						
						
							| Preamp: | 
							0 | 
						
						
							| Current location: | 
							Bochum | 
						
						
							|   | 
							  | 
						
						
							| Installation information: | 
							  | 
						
						
							| Label: | 
							red | 
						
						
							|   | 
							  | 
						
						
							| Manufacturer information: | 
							  | 
						
						
							| Wafer position: | 
							C11 | 
						
						
							| Break-through voltage: | 
							409 V | 
						
						
							| Voltage for Gain 100 (T=+25°C): | 
							380.7 V | 
						
						
							| Dark current: | 
							9.2 nA | 
						
						
							|   | 
							  | 
						
						
							| Screening Logistics: | 
							  | 
						
						
							| Available: | 
							No | 
						
						
							| Storage Box: | 
							74 | 
						
						
							| Position in Box: | 
							14 | 
						
						
							| EP1 batch: | 
							55 | 
						
						
							| EP1 batch after irradiation: | 
							10171 | 
						
						
							|   | 
							  | 
						
						
							| Shipment: | 
							  | 
						
						
							| Grid number: | 
							none | 
						
						
							| Position in grid: | 
							none | 
						
						
							| Arrival for irradiation: | 
							26. Aug 2019 | 
						
						
							| Sent for analysis after irradiation: | 
							02. Sep 2019 | 
						
						
							| Return for assembly: | 
							none | 
						
						
							|   | 
							  | 
						
						
							| Irradiation: | 
							  | 
						
						
							| Date: | 
							28. Aug 2019 | 
						
						
							| Dose used: | 
							37 Gy | 
						
						
							| Temperature: | 
							20 °C | 
						
						
							| Position: | 
							none | 
						
						
							| Bias voltage: | 
							none      | 
						
						
							|   | 
							  | 
						
						
							| Annealing: | 
							  | 
						
						
							| Date: | 
							31. Aug 2019 | 
						
						
							| Temperature: | 
							80 °C | 
						
						
							| Duration: | 
							48 h | 
						
						
							|   | 
							  | 
						
						
							| Measurement results: | 
							  | 
						
						
							| Voltage for Gain 100: | 
							T = +20 °C: 380.990565 V     T = -25 °C: 345.897617 V | 
						
						
							| Voltage for Gain 150: | 
							T = +20 °C: 388.8169979 V     T = -25 °C: 353.2634227 V | 
						
						
							| Voltage for Gain 200: | 
							T = +20 °C: 393.1845123 V     T = -25 °C: 357.4044517 V | 
						
						
							| Gain/Voltage slope at M = 100: | 
							T = +20 °C: 4.608868672 V-1     T = -25 °C: 4.969734925 V-1 | 
						
						
							| Gain/Voltage slope at M = 150: | 
							T = +20 °C: 9.12785802 V-1     T = -25 °C: 9.190349267 V-1 | 
						
						
							| Gain/Voltage slope at M = 200: | 
							T = +20 °C: 14.16295531 V-1     T = -25 °C: 14.52559552 V-1 | 
						
						
							| Break-through voltage: | 
							T = +20 °C: 408.7577712 V     T = -25 °C: 372.4829686 V | 
						
						
							|   | 
							  | 
						
						
							| Notes: | 
							
								
							 | 
						
					
					
						
					
					Characteristics
					
						
							| Temperature | 
							Measurement | 
							Notes | 
						
						
							| No characteristics available! | 
						
					
					
						
						Upload new characteristic...
						
					
					Progression of the current during irradiation
					
						
							| Upload time | 
							Notes | 
						
						
							| No data available! | 
						
					
					
						
						Upload new progression data...
						
					
					Version history