Subdetector specification:  
Serial: 1410015849
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: G12
Break-through voltage: 427 V
Voltage for Gain 100 (T=+25°C): 398.8 V
Dark current: 8.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 37
Position in Box: 10
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: 391
Position in grid: 12
Arrival for irradiation: 29. Sep 2017
Sent for analysis after irradiation: 03. Jan 2019
Return for assembly: none
   
Irradiation:  
Date: 05. Dec 2018
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 10. Dec 2018
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 399.2395305 V     T = -25 °C: 362.5275733 V
Voltage for Gain 150: T = +20 °C: 407.1758292 V     T = -25 °C: 370.3389769 V
Voltage for Gain 200: T = +20 °C: 411.5839513 V     T = -25 °C: 374.7125098 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.387263089 V-1     T = -25 °C: 4.700649231 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.19056945 V-1     T = -25 °C: 9.164406731 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.03285901 V-1     T = -25 °C: 13.95956307 V-1
Break-through voltage: T = +20 °C: 426.9075603 V     T = -25 °C: 390.2223698 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:52:47 CET jreher Updated box and position via API.
03. Feb 2021 18:18:42 CET jreher Updated box and position via API.
11. Aug 2020 17:25:40 CEST jreher Updated APD characteristics via API.
28. Nov 2019 12:23:11 CET jreher Updated APD location via API.
25. Nov 2019 17:13:47 CET jreher Updated box and position via API.
13. Jun 2019 11:29:55 CEST jreher Updated APD location via API.
12. Jun 2019 16:48:09 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:33:05 CEST jreher Updated APD batch assignment via API.
03. Jan 2019 13:01:00 CET markus.moritz
29. Sep 2017 12:25:44 CEST markus.moritz Entered arrival for irradiation date.
01. Aug 2016 17:29:25 CEST Tobias Assigned to grid via information extracted from APD DB.
23. Jun 2016 10:39:42 CEST Tobias Imported from Hamamatsu datasheet.