Show Hamamatsu Avalanche Photo Diode 1206013085 - Archived data from Wed, 03. February 2021 18:19:44 CET
This is all the information about APD 1206013085. If it is wrong, edit the data.
| Subdetector specification: | |
|---|---|
| Serial: | 1206013085 |
| Type: | Hamamatsu Avalanche Photo Diode |
| Detector: | unassigned |
| Unit: | unassigned |
| Preamp: | 0 |
| Current location: | In transit to Mainz |
| Installation information: | |
| Label: | none |
| Manufacturer information: | |
| Wafer position: | C04 |
| Break-through voltage: | 436 V |
| Voltage for Gain 100 (T=+25°C): | 407.1 V |
| Dark current: | 9.8 nA |
| Screening Logistics: | |
| Available: | Yes |
| Storage Box: | 45 |
| Position in Box: | 4 |
| EP1 batch: | none |
| EP1 batch after irradiation: | none |
| Shipment: | |
| Grid number: | 267 |
| Position in grid: | 18 |
| Arrival for irradiation: | 11. Jul 2016 |
| Sent for analysis after irradiation: | 15. Jul 2016 |
| Return for assembly: | 19. Sep 2016 |
| Irradiation: | |
| Date: | 13. Jul 2016 |
| Dose used: | 30 Gy |
| Temperature: | 25 °C |
| Position: | 1 |
| Bias voltage: | 320 V (connection unknown) |
| Annealing: | |
| Date: | 13. Jul 2016 |
| Temperature: | 80 °C |
| Duration: | 10 h |
| Measurement results: | |
| Voltage for Gain 100: | T = +20 °C: 407.9857656 V T = -25 °C: 370.892831 V |
| Voltage for Gain 150: | T = +20 °C: 415.9277422 V T = -25 °C: 378.7727314 V |
| Voltage for Gain 200: | T = +20 °C: 420.3406343 V T = -25 °C: 383.1502147 V |
| Gain/Voltage slope at M = 100: | T = +20 °C: 4.679464239 V-1 T = -25 °C: 4.72337423 V-1 |
| Gain/Voltage slope at M = 150: | T = +20 °C: 9.049803174 V-1 T = -25 °C: 9.121527478 V-1 |
| Gain/Voltage slope at M = 200: | T = +20 °C: 14.91438857 V-1 T = -25 °C: 14.3069594 V-1 |
| Break-through voltage: | T = +20 °C: 436.1061582 V T = -25 °C: 399.0503446 V |
| Notes: |
Version history
| Time | Author | Change comment |
|---|---|---|
| 05. Feb 2021 10:53:56 CET | jreher | Updated box and position via API. |
| 03. Feb 2021 18:19:44 CET | jreher | Updated box and position via API. |
| 07. Oct 2020 15:46:03 CEST | jreher | Updated APD location via API. |
| 11. Aug 2020 17:26:57 CEST | jreher | Updated APD characteristics via API. |
| 27. Nov 2019 19:36:58 CET | jreher | Updated APD location via API. |
| 25. Nov 2019 17:14:28 CET | jreher | Updated box and position via API. |
| 04. Jul 2019 11:11:43 CEST | jreher | Updated APD location via API. |
| 13. Jun 2019 10:01:32 CEST | jreher | Updated APD batch assignment via API. |
| 23. Sep 2016 09:05:44 CEST | markus.moritz | Entered return for assembly date. |
| 03. Aug 2016 16:36:53 CEST | markus.moritz | Entered irradiation information. |
| 02. Aug 2016 15:00:12 CEST | christopher.l.hahn | Entered annealing information. |
| 02. Aug 2016 10:44:37 CEST | markus.moritz | Entered shipped after irradiation date. |
| 02. Aug 2016 10:43:54 CEST | markus.moritz | Entered arrival for irradiation date. |
| 01. Aug 2016 17:28:02 CEST | Tobias | Assigned to grid via information extracted from APD DB. |
| 22. Jun 2016 15:39:31 CEST | Tobias | Imported from Hamamatsu datasheet. |