RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 0911009805 - Archived data from Tue, 11. August 2020 17:40:46 CEST


This is all the information about APD 0911009805. If it is wrong, edit the data.

Subdetector specification:  
Serial: 0911009805
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D03
Break-through voltage: 405 V
Voltage for Gain 100 (T=+25°C): 376.4 V
Dark current: 3.2 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 292
Position in Box: 45
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: 210
Position in grid: 17
Arrival for irradiation: 10. Jun 2016
Sent for analysis after irradiation: 23. Jun 2016
Return for assembly: 04. Aug 2016
   
Irradiation:  
Date: 20. Jun 2016
Dose used: 37 Gy
Temperature: 25 °C
Position: 2
Bias voltage: 320 V     (connection unknown)
   
Annealing:  
Date: 20. Jun 2016
Temperature: 80 °C
Duration: 8 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 376.7546473 V     T = -25 °C: 341.9762398 V
Voltage for Gain 150: T = +20 °C: 384.6485548 V     T = -25 °C: 349.4188655 V
Voltage for Gain 200: T = +20 °C: 389.026599 V     T = -25 °C: 353.6224724 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.681195683 V-1     T = -25 °C: 4.947046104 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.317636716 V-1     T = -25 °C: 9.039789383 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.48748753 V-1     T = -25 °C: 15.95269918 V-1
Break-through voltage: T = +20 °C: 406.0956262 V     T = -25 °C: 368.4439082 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:26:01 CET jreher Updated box and position via API.
07. Oct 2020 15:45:00 CEST jreher Updated APD location via API.
11. Aug 2020 17:40:46 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:41:03 CEST jreher Updated box and position via API.
27. Nov 2019 20:19:14 CET jreher Updated APD location via API.
25. Nov 2019 17:31:39 CET jreher Updated box and position via API.
22. Oct 2019 14:33:22 CEST spankonin Updated APD location via API.
15. Oct 2019 10:59:39 CEST jgrochowski Updated APD batch assignment via API.
04. Aug 2016 13:25:45 CEST markus.moritz Entered return for assembly date.
03. Aug 2016 16:12:47 CEST markus.moritz Entered shipped after irradiation date.
03. Aug 2016 16:12:38 CEST markus.moritz Entered arrival for irradiation date.
02. Aug 2016 16:28:36 CEST markus.moritz Entered irradiation information.
02. Aug 2016 15:21:42 CEST christopher.l.hahn Entered annealing information.
01. Aug 2016 17:27:39 CEST Tobias Assigned to grid via information extracted from APD DB.
22. Jun 2016 15:34:22 CEST Tobias Imported from Hamamatsu datasheet.