RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 0617005279 - Archived data from Wed, 07. October 2020 15:44:10 CEST


This is all the information about APD 0617005279. If it is wrong, edit the data.

Subdetector specification:  
Serial: 0617005279
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: In transit to Mainz
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: H10
Break-through voltage: 451 V
Voltage for Gain 100 (T=+25°C): 423.3 V
Dark current: 15.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 246
Position in Box: 49
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 423.9536742 V     T = -25 °C: 387.4948633 V
Voltage for Gain 150: T = +20 °C: 431.790607 V     T = -25 °C: 395.2393046 V
Voltage for Gain 200: T = +20 °C: 436.1177489 V     T = -25 °C: 399.5238686 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.59482363 V-1     T = -25 °C: 4.486896313 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.204384106 V-1     T = -25 °C: 8.860937542 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.39950999 V-1     T = -25 °C: 15.71970072 V-1
Break-through voltage: T = +20 °C: 439.9849965 V     T = -25 °C: 414.03692 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:20:00 CET jreher Updated box and position via API.
07. Oct 2020 15:44:10 CEST jreher Updated APD location via API.
11. Aug 2020 17:48:15 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:31:52 CEST jreher Updated box and position via API.
27. Nov 2019 20:12:32 CET jreher Updated APD location via API.
25. Nov 2019 17:33:21 CET jreher Updated box and position via API.
21. Nov 2019 16:52:53 CET spankonin Updated APD location via API.
21. Nov 2019 16:44:12 CET spankonin Updated APD batch assignment via API.
18. Jun 2019 10:18:41 CEST Tobias Imported from Hamamatsu datasheet.