RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3315036362 - Archived data from Tue, 11. August 2020 18:02:46 CEST


This is all the information about APD 3315036362. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3315036362
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D05
Break-through voltage: 406 V
Voltage for Gain 100 (T=+25°C): 377.2 V
Dark current: 17.6 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 155
Position in Box: 18
EP1 batch: 184
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 377.6134967 V     T = -25 °C: 343.5338468 V
Voltage for Gain 150: T = +20 °C: 385.2632216 V     T = -25 °C: 350.6153137 V
Voltage for Gain 200: T = +20 °C: 389.553734 V     T = -25 °C: 354.58774 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.562066741 V-1     T = -25 °C: 4.931958056 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.951867439 V-1     T = -25 °C: 10.32941947 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.63552454 V-1     T = -25 °C: 16.83020417 V-1
Break-through voltage: T = +20 °C: 404.9441087 V     T = -25 °C: 369.2235119 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:08:14 CET jreher Updated box and position via API.
11. Aug 2020 18:02:46 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:52:23 CEST jreher Updated APD location via API.
03. Feb 2020 14:22:09 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:49:30 CET jreher Updated APD location via API.
27. Nov 2019 19:56:15 CET jreher Updated APD location via API.
25. Nov 2019 17:23:09 CET jreher Updated box and position via API.
01. Oct 2019 11:33:05 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:56:49 CEST jreher Updated APD location via API.
03. Jan 2019 00:15:26 CET Tobias Corrected dark current unit.
19. Sep 2018 14:05:34 CEST Tobias Imported from Hamamatsu datasheet.