RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3312036147 - Archived data from Tue, 11. August 2020 18:02:15 CEST


This is all the information about APD 3312036147. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3312036147
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: H08
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 384.4 V
Dark current: 30.7 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 152
Position in Box: 12
EP1 batch: 179
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 384.4430048 V     T = -25 °C: 349.4017197 V
Voltage for Gain 150: T = +20 °C: 392.215098 V     T = -25 °C: 356.6934131 V
Voltage for Gain 200: T = +20 °C: 396.5520974 V     T = -25 °C: 360.7812393 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.506279169 V-1     T = -25 °C: 4.86389179 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.867936537 V-1     T = -25 °C: 9.965257545 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.58395887 V-1     T = -25 °C: 16.03962546 V-1
Break-through voltage: T = +20 °C: 411.0555723 V     T = -25 °C: 375.1871192 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:07:50 CET jreher Updated box and position via API.
11. Aug 2020 18:02:15 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:52:08 CEST jreher Updated APD location via API.
03. Feb 2020 14:24:04 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:49:18 CET jreher Updated APD location via API.
27. Nov 2019 19:55:41 CET jreher Updated APD location via API.
25. Nov 2019 17:22:53 CET jreher Updated box and position via API.
01. Oct 2019 11:29:20 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:56:24 CEST jreher Updated APD location via API.
03. Jan 2019 00:15:21 CET Tobias Corrected dark current unit.
19. Sep 2018 14:05:31 CEST Tobias Imported from Hamamatsu datasheet.