RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3206034964 - Archived data from Tue, 11. August 2020 18:08:26 CEST


This is all the information about APD 3206034964. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3206034964
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C11
Break-through voltage: 411 V
Voltage for Gain 100 (T=+25°C): 382.6 V
Dark current: 21.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 136
Position in Box: 18
EP1 batch: 155
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 408.9779769 V     T = -25 °C: 365.7849753 V
Voltage for Gain 150: T = +20 °C: 408.9834393 V     T = -25 °C: 368.3641707 V
Voltage for Gain 200: T = +20 °C: 408.9889017 V     T = -25 °C: 369.8670539 V
Gain/Voltage slope at M = 100: T = +20 °C: 9153.442351 V-1     T = -25 °C: 11.47428426 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9153.442351 V-1     T = -25 °C: 28.95667284 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 9153.442351 V-1     T = -25 °C: 36.23367296 V-1
Break-through voltage: T = +20 °C: 410.0158131 V     T = -25 °C: 373.8562262 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:05:51 CET jreher Updated box and position via API.
11. Aug 2020 18:08:26 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:51:00 CEST jreher Updated APD location via API.
09. Mar 2020 17:39:25 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:48:21 CET jreher Updated APD location via API.
27. Nov 2019 19:52:49 CET jreher Updated APD location via API.
25. Nov 2019 17:21:41 CET jreher Updated box and position via API.
20. Sep 2019 18:44:31 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:54:25 CEST jreher Updated APD location via API.
03. Jan 2019 00:14:55 CET Tobias Corrected dark current unit.
14. Aug 2018 16:04:19 CEST Tobias Imported from Hamamatsu datasheet.