RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3004033447 - Archived data from Tue, 11. August 2020 18:06:41 CEST


This is all the information about APD 3004033447. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3004033447
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: B13
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 386.6 V
Dark current: 27.4 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 114
Position in Box: 30
EP1 batch: 120
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 386.4232676 V     T = -25 °C: 350.3607577 V
Voltage for Gain 150: T = +20 °C: 394.2763813 V     T = -25 °C: 357.9284677 V
Voltage for Gain 200: T = +20 °C: 398.6379776 V     T = -25 °C: 362.1736911 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.416358853 V-1     T = -25 °C: 4.877057705 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.687270191 V-1     T = -25 °C: 9.435748908 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.04845888 V-1     T = -25 °C: 14.09494079 V-1
Break-through voltage: T = +20 °C: 411.0093035 V     T = -25 °C: 377.9303193 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:03:06 CET jreher Updated box and position via API.
11. Aug 2020 18:06:41 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:49:23 CEST jreher Updated APD location via API.
25. Feb 2020 16:44:39 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:47:02 CET jreher Updated APD location via API.
27. Nov 2019 19:48:49 CET jreher Updated APD location via API.
25. Nov 2019 17:19:58 CET jreher Updated box and position via API.
27. Aug 2019 12:10:39 CEST jgrochowski Updated APD batch assignment via API.
05. Jul 2019 13:51:34 CEST jreher Updated APD location via API.
03. Jan 2019 00:14:21 CET Tobias Corrected dark current unit.
06. Jul 2018 22:38:47 CEST Tobias Imported from Hamamatsu datasheet.