Show Hamamatsu Avalanche Photo Diode 0803007417 - Archived data from Wed, 03. February 2021 18:20:40 CET
					This is all the information about APD 0803007417. If it is wrong, edit the data.
| Subdetector specification: | |
|---|---|
| Serial: | 0803007417 | 
| Type: | Hamamatsu Avalanche Photo Diode | 
| Detector: | unassigned | 
| Unit: | unassigned | 
| Preamp: | 0 | 
| Current location: | Bochum | 
| Installation information: | |
| Label: | none | 
| Manufacturer information: | |
| Wafer position: | E05 | 
| Break-through voltage: | 406 V | 
| Voltage for Gain 100 (T=+25°C): | 377.9 V | 
| Dark current: | 3.5 nA | 
| Screening Logistics: | |
| Available: | Yes | 
| Storage Box: | 52 | 
| Position in Box: | 33 | 
| EP1 batch: | none | 
| EP1 batch after irradiation: | none | 
| Shipment: | |
| Grid number: | none | 
| Position in grid: | none | 
| Arrival for irradiation: | 18. Apr 2019 | 
| Sent for analysis after irradiation: | 04. Jun 2019 | 
| Return for assembly: | none | 
| Irradiation: | |
| Date: | 29. May 2019 | 
| Dose used: | 37 Gy | 
| Temperature: | 20 °C | 
| Position: | none | 
| Bias voltage: | none | 
| Annealing: | |
| Date: | 29. May 2019 | 
| Temperature: | 80 °C | 
| Duration: | 48 h | 
| Measurement results: | |
| Voltage for Gain 100: | T = +20 °C: 377.8636091 V T = -25 °C: 342.2829455 V | 
| Voltage for Gain 150: | T = +20 °C: 385.705674 V T = -25 °C: 349.6234995 V | 
| Voltage for Gain 200: | T = +20 °C: 390.0670467 V T = -25 °C: 353.7628155 V | 
| Gain/Voltage slope at M = 100: | T = +20 °C: 4.532088053 V-1 T = -25 °C: 4.693944168 V-1 | 
| Gain/Voltage slope at M = 150: | T = +20 °C: 8.772419882 V-1 T = -25 °C: 9.335978855 V-1 | 
| Gain/Voltage slope at M = 200: | T = +20 °C: 14.89563527 V-1 T = -25 °C: 16.26951099 V-1 | 
| Break-through voltage: | T = +20 °C: 405.7481591 V T = -25 °C: 368.9410222 V | 
| Notes: | 
						
					
Version history
| Time | Author | Change comment | 
|---|---|---|
| 21. Apr 2021 14:48:26 CEST | oafedulidis | Updated box and position via API. | 
| 05. Feb 2021 10:55:05 CET | jreher | Updated box and position via API. | 
| 03. Feb 2021 18:20:40 CET | jreher | Updated box and position via API. | 
| 11. Aug 2020 17:27:35 CEST | jreher | Updated APD characteristics via API. | 
| 27. Nov 2019 19:37:55 CET | jreher | Updated APD location via API. | 
| 25. Nov 2019 17:15:03 CET | jreher | Updated box and position via API. | 
| 04. Jul 2019 11:12:20 CEST | jreher | Updated APD location via API. | 
| 18. Jun 2019 13:12:27 CEST | jgrochowski | Updated APD batch assignment via API. | 
| 04. Jun 2019 15:51:29 CEST | markus.moritz | |
| 22. Jun 2016 15:22:36 CEST | Tobias | Imported from Hamamatsu datasheet. | 
 
						 
					