RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2804031093 - Archived data from Wed, 28. April 2021 13:27:29 CEST


This is all the information about APD 2804031093. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2804031093
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E07
Break-through voltage: 406 V
Voltage for Gain 100 (T=+25°C): 376.8 V
Dark current: 7.7 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 60
Position in Box: 47
EP1 batch: 74
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 377.2245587 V     T = -25 °C: 342.1279768 V
Voltage for Gain 150: T = +20 °C: 385.0477799 V     T = -25 °C: 349.4472366 V
Voltage for Gain 200: T = +20 °C: 389.4223976 V     T = -25 °C: 353.5489959 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.571865981 V-1     T = -25 °C: 4.913842525 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.940717204 V-1     T = -25 °C: 9.129672116 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.58419108 V-1     T = -25 °C: 16.45594839 V-1
Break-through voltage: T = +20 °C: 405.4487345 V     T = -25 °C: 369.0664036 V
   
Notes:


Version history

Time Author Change comment
26. Oct 2021 10:25:35 CEST oafedulidis Assigned to capsule by script
28. Apr 2021 13:27:29 CEST oafedulidis Updated box and position via API.
26. Apr 2021 15:22:14 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:59:20 CET jreher Updated box and position via API.
03. Feb 2021 18:24:50 CET jreher Updated box and position via API.
11. Aug 2020 17:39:06 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:43:56 CET jreher Updated APD location via API.
25. Nov 2019 17:17:40 CET jreher Updated box and position via API.
13. Sep 2019 10:29:15 CEST jgrochowski Updated APD batch assignment via API.
04. Sep 2019 13:27:33 CEST markus.moritz
25. Jul 2019 12:12:00 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:47:36 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:34 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:05:38 CEST Tobias Imported from Hamamatsu datasheet.