Subdetector specification:  
Serial: 2802030931
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C07
Break-through voltage: 406 V
Voltage for Gain 100 (T=+25°C): 379 V
Dark current: 13.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 64
Position in Box: 37
EP1 batch: 71
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 379.4750069 V     T = -25 °C: 344.4969026 V
Voltage for Gain 150: T = +20 °C: 387.2723421 V     T = -25 °C: 351.8263629 V
Voltage for Gain 200: T = +20 °C: 391.6174878 V     T = -25 °C: 355.9286985 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.484505165 V-1     T = -25 °C: 4.769986417 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.761219301 V-1     T = -25 °C: 9.737377927 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.2920968 V-1     T = -25 °C: 15.52323279 V-1
Break-through voltage: T = +20 °C: 406.2237807 V     T = -25 °C: 370.0492178 V
   
Notes:


Version history

Time Author Change comment
16. Nov 2021 09:41:50 CET oafedulidis Assigned to capsule by script
16. Jun 2021 16:02:37 CEST oafedulidis Updated box and position via API.
16. Jun 2021 15:58:55 CEST oafedulidis Updated box and position via API.
16. Jun 2021 15:51:44 CEST oafedulidis Updated box and position via API.
03. May 2021 10:25:07 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:59:10 CET jreher Updated box and position via API.
03. Feb 2021 18:24:40 CET jreher Updated box and position via API.
11. Aug 2020 17:38:54 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:43:43 CET jreher Updated APD location via API.
25. Nov 2019 17:17:34 CET jreher Updated box and position via API.
13. Sep 2019 10:28:00 CEST jgrochowski Updated APD batch assignment via API.
04. Sep 2019 13:03:00 CEST markus.moritz
25. Jul 2019 12:12:46 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:47:26 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:32 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:05:09 CEST Tobias Imported from Hamamatsu datasheet.