Show Hamamatsu Avalanche Photo Diode 2707030541 - Archived data from Tue, 11. August 2020 17:38:05 CEST
This is all the information about APD 2707030541. If it is wrong, edit the data.
| Subdetector specification: | |
|---|---|
| Serial: | 2707030541 |
| Type: | Hamamatsu Avalanche Photo Diode |
| Detector: | unassigned |
| Unit: | unassigned |
| Preamp: | 0 |
| Current location: | Bochum |
| Installation information: | |
| Label: | none |
| Manufacturer information: | |
| Wafer position: | C09 |
| Break-through voltage: | 409 V |
| Voltage for Gain 100 (T=+25°C): | 380.5 V |
| Dark current: | 6.2 nA |
| Screening Logistics: | |
| Available: | Yes |
| Storage Box: | 93 |
| Position in Box: | 35 |
| EP1 batch: | 63 |
| EP1 batch after irradiation: | none |
| Shipment: | |
| Grid number: | none |
| Position in grid: | none |
| Arrival for irradiation: | 26. Aug 2019 |
| Sent for analysis after irradiation: | 02. Sep 2019 |
| Return for assembly: | none |
| Irradiation: | |
| Date: | 28. Aug 2019 |
| Dose used: | 37 Gy |
| Temperature: | 20 °C |
| Position: | none |
| Bias voltage: | none |
| Annealing: | |
| Date: | 31. Aug 2019 |
| Temperature: | 80 °C |
| Duration: | 48 h |
| Measurement results: | |
| Voltage for Gain 100: | T = +20 °C: 381.159566 V T = -25 °C: 345.6884234 V |
| Voltage for Gain 150: | T = +20 °C: 388.9990925 V T = -25 °C: 353.0770788 V |
| Voltage for Gain 200: | T = +20 °C: 393.3677582 V T = -25 °C: 357.2222012 V |
| Gain/Voltage slope at M = 100: | T = +20 °C: 4.559008192 V-1 T = -25 °C: 4.690208859 V-1 |
| Gain/Voltage slope at M = 150: | T = +20 °C: 9.034044988 V-1 T = -25 °C: 9.478352382 V-1 |
| Gain/Voltage slope at M = 200: | T = +20 °C: 14.05811377 V-1 T = -25 °C: 15.00523435 V-1 |
| Break-through voltage: | T = +20 °C: 408.9066431 V T = -25 °C: 372.5151011 V |
| Notes: |
Version history
| Time | Author | Change comment |
|---|---|---|
| 03. May 2021 10:26:08 CEST | oafedulidis | Updated box and position via API. |
| 05. Feb 2021 10:58:30 CET | jreher | Updated box and position via API. |
| 03. Feb 2021 18:24:00 CET | jreher | Updated box and position via API. |
| 11. Aug 2020 17:38:05 CEST | jreher | Updated APD characteristics via API. |
| 27. Nov 2019 19:42:48 CET | jreher | Updated APD location via API. |
| 25. Nov 2019 17:17:10 CET | jreher | Updated box and position via API. |
| 10. Sep 2019 11:19:58 CEST | jgrochowski | Updated APD batch assignment via API. |
| 04. Sep 2019 10:23:15 CEST | markus.moritz | |
| 25. Jul 2019 12:14:41 CEST | spankonin | Updated APD batch assignment via API. |
| 05. Jul 2019 13:46:46 CEST | jreher | Updated APD location via API. |
| 03. Jan 2019 00:13:23 CET | Tobias | Corrected dark current unit. |
| 24. May 2018 14:39:23 CEST | Tobias | Updated location to "Bochum". |
| 18. May 2018 12:03:30 CEST | Tobias | Imported from Hamamatsu datasheet. |