Show Hamamatsu Avalanche Photo Diode 2705030370 - Archived data from Wed, 03. February 2021 18:23:45 CET
This is all the information about APD 2705030370. If it is wrong, edit the data.
| Subdetector specification: | |
|---|---|
| Serial: | 2705030370 |
| Type: | Hamamatsu Avalanche Photo Diode |
| Detector: | unassigned |
| Unit: | unassigned |
| Preamp: | 0 |
| Current location: | Bochum |
| Installation information: | |
| Label: | none |
| Manufacturer information: | |
| Wafer position: | A07 |
| Break-through voltage: | 416 V |
| Voltage for Gain 100 (T=+25°C): | 387.7 V |
| Dark current: | 7.4 nA |
| Screening Logistics: | |
| Available: | Yes |
| Storage Box: | 77 |
| Position in Box: | 23 |
| EP1 batch: | 60 |
| EP1 batch after irradiation: | none |
| Shipment: | |
| Grid number: | none |
| Position in grid: | none |
| Arrival for irradiation: | 26. Aug 2019 |
| Sent for analysis after irradiation: | 02. Sep 2019 |
| Return for assembly: | none |
| Irradiation: | |
| Date: | 28. Aug 2019 |
| Dose used: | 37 Gy |
| Temperature: | 20 °C |
| Position: | none |
| Bias voltage: | none |
| Annealing: | |
| Date: | 31. Aug 2019 |
| Temperature: | 80 °C |
| Duration: | 48 h |
| Measurement results: | |
| Voltage for Gain 100: | T = +20 °C: 387.9143791 V T = -25 °C: 351.9006521 V |
| Voltage for Gain 150: | T = +20 °C: 395.8539794 V T = -25 °C: 359.4580944 V |
| Voltage for Gain 200: | T = +20 °C: 400.2712453 V T = -25 °C: 363.704885 V |
| Gain/Voltage slope at M = 100: | T = +20 °C: 4.579468584 V-1 T = -25 °C: 4.868545004 V-1 |
| Gain/Voltage slope at M = 150: | T = +20 °C: 8.994933295 V-1 T = -25 °C: 8.859717422 V-1 |
| Gain/Voltage slope at M = 200: | T = +20 °C: 13.93445702 V-1 T = -25 °C: 15.42801789 V-1 |
| Break-through voltage: | T = +20 °C: 415.2300969 V T = -25 °C: 378.8348328 V |
| Notes: |
Version history
| Time | Author | Change comment |
|---|---|---|
| 05. Feb 2021 10:58:13 CET | jreher | Updated box and position via API. |
| 03. Feb 2021 18:23:45 CET | jreher | Updated box and position via API. |
| 11. Aug 2020 17:37:51 CEST | jreher | Updated APD characteristics via API. |
| 27. Nov 2019 19:42:26 CET | jreher | Updated APD location via API. |
| 25. Nov 2019 17:17:01 CET | jreher | Updated box and position via API. |
| 06. Sep 2019 13:13:15 CEST | spankonin | Updated APD batch assignment via API. |
| 03. Sep 2019 14:34:26 CEST | markus.moritz | |
| 08. Jul 2019 14:56:12 CEST | spankonin | Updated APD batch assignment via API. |
| 05. Jul 2019 13:46:30 CEST | jreher | Updated APD location via API. |
| 03. Jan 2019 00:13:19 CET | Tobias | Corrected dark current unit. |
| 24. May 2018 14:39:23 CEST | Tobias | Updated location to "Bochum". |
| 18. May 2018 12:02:51 CEST | Tobias | Imported from Hamamatsu datasheet. |