RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2701030036 - Archived data from Fri, 05. February 2021 10:57:43 CET


This is all the information about APD 2701030036. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2701030036
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: B07
Break-through voltage: 414 V
Voltage for Gain 100 (T=+25°C): 386.1 V
Dark current: 9.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 73
Position in Box: 37
EP1 batch: 54
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 28. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 386.6874908 V     T = -25 °C: 350.8967079 V
Voltage for Gain 150: T = +20 °C: 394.4784656 V     T = -25 °C: 358.2939243 V
Voltage for Gain 200: T = +20 °C: 398.7908719 V     T = -25 °C: 362.4475239 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.407650483 V-1     T = -25 °C: 4.959810974 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.642707217 V-1     T = -25 °C: 9.203849726 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.08045243 V-1     T = -25 °C: 16.43123617 V-1
Break-through voltage: T = +20 °C: 413.3144659 V     T = -25 °C: 376.7372964 V
   
Notes:


Version history

Time Author Change comment
10. Jun 2022 12:31:07 CEST oafedulidis Assigned to capsule by script
27. Aug 2021 11:14:28 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:57:43 CET jreher Updated box and position via API.
03. Feb 2021 18:23:16 CET jreher Updated box and position via API.
11. Aug 2020 17:37:16 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:41:42 CET jreher Updated APD location via API.
25. Nov 2019 17:16:43 CET jreher Updated box and position via API.
06. Sep 2019 13:10:23 CEST spankonin Updated APD batch assignment via API.
02. Sep 2019 14:47:10 CEST markus.moritz
08. Jul 2019 14:54:23 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:45:59 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:13 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:01:48 CEST Tobias Imported from Hamamatsu datasheet.