RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2701030026 - Archived data from Tue, 11. August 2020 17:37:14 CEST


This is all the information about APD 2701030026. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2701030026
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2701030026/2502027677
Unit: #947 (barcode 1309010714)
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: blue
   
Manufacturer information:  
Wafer position: G08
Break-through voltage: 411 V
Voltage for Gain 100 (T=+25°C): 383.6 V
Dark current: 7 nA
   
Screening Logistics:  
Available: No
Storage Box: 73
Position in Box: 27
EP1 batch: 54
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 383.3074808 V     T = -25 °C: 347.6698318 V
Voltage for Gain 150: T = +20 °C: 391.1114727 V     T = -25 °C: 355.0929148 V
Voltage for Gain 200: T = +20 °C: 395.4486097 V     T = -25 °C: 359.2557997 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.496293385 V-1     T = -25 °C: 4.990375787 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.846655734 V-1     T = -25 °C: 9.271503556 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.54570669 V-1     T = -25 °C: 14.65103817 V-1
Break-through voltage: T = +20 °C: 410.3868467 V     T = -25 °C: 374.1549524 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:57:42 CET jreher Updated box and position via API.
03. Feb 2021 18:23:15 CET jreher Updated box and position via API.
11. Aug 2020 17:37:14 CEST jreher Updated APD characteristics via API.
31. Mar 2020 08:53:00 CEST jreher Assigned to capsule by script
13. Feb 2020 14:23:01 CET jreher Updated box and position via API.
27. Nov 2019 19:41:42 CET jreher Updated APD location via API.
25. Nov 2019 17:16:43 CET jreher Updated box and position via API.
06. Sep 2019 13:10:03 CEST spankonin Updated APD batch assignment via API.
02. Sep 2019 14:26:32 CEST markus.moritz
08. Jul 2019 14:54:23 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:45:58 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:12 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:01:47 CEST Tobias Imported from Hamamatsu datasheet.