Show Hamamatsu Avalanche Photo Diode 2610029910 - Archived data from Mon, 10. May 2021 14:01:17 CEST
This is all the information about APD 2610029910. If it is wrong, edit the data.
| Subdetector specification: | |
|---|---|
| Serial: | 2610029910 |
| Type: | Hamamatsu Avalanche Photo Diode |
| Detector: | unassigned |
| Unit: | unassigned |
| Preamp: | 0 |
| Current location: | Bochum |
| Installation information: | |
| Label: | none |
| Manufacturer information: | |
| Wafer position: | B05 |
| Break-through voltage: | 412 V |
| Voltage for Gain 100 (T=+25°C): | 383.3 V |
| Dark current: | 6.6 nA |
| Screening Logistics: | |
| Available: | Yes |
| Storage Box: | 72 |
| Position in Box: | 13 |
| EP1 batch: | 52 |
| EP1 batch after irradiation: | none |
| Shipment: | |
| Grid number: | none |
| Position in grid: | none |
| Arrival for irradiation: | 26. Aug 2019 |
| Sent for analysis after irradiation: | 02. Sep 2019 |
| Return for assembly: | none |
| Irradiation: | |
| Date: | 28. Aug 2019 |
| Dose used: | 37 Gy |
| Temperature: | 20 °C |
| Position: | none |
| Bias voltage: | none |
| Annealing: | |
| Date: | 31. Aug 2019 |
| Temperature: | 80 °C |
| Duration: | 48 h |
| Measurement results: | |
| Voltage for Gain 100: | T = +20 °C: 383.1658825 V T = -25 °C: 347.5688058 V |
| Voltage for Gain 150: | T = +20 °C: 390.9832035 V T = -25 °C: 355.0041517 V |
| Voltage for Gain 200: | T = +20 °C: 395.3332244 V T = -25 °C: 359.1808559 V |
| Gain/Voltage slope at M = 100: | T = +20 °C: 4.579057543 V-1 T = -25 °C: 4.706374156 V-1 |
| Gain/Voltage slope at M = 150: | T = +20 °C: 9.024205298 V-1 T = -25 °C: 9.447117903 V-1 |
| Gain/Voltage slope at M = 200: | T = +20 °C: 14.12031745 V-1 T = -25 °C: 14.84246289 V-1 |
| Break-through voltage: | T = +20 °C: 410.7580905 V T = -25 °C: 374.657871 V |
| Notes: |
Version history
| Time | Author | Change comment |
|---|---|---|
| 25. Sep 2021 16:11:26 CEST | oafedulidis | Assigned to capsule by script |
| 10. May 2021 14:01:17 CEST | oafedulidis | Updated box and position via API. |
| 05. Feb 2021 10:57:33 CET | jreher | Updated box and position via API. |
| 03. Feb 2021 18:23:06 CET | jreher | Updated box and position via API. |
| 11. Aug 2020 17:37:05 CEST | jreher | Updated APD characteristics via API. |
| 27. Nov 2019 19:41:29 CET | jreher | Updated APD location via API. |
| 25. Nov 2019 17:16:36 CET | jreher | Updated box and position via API. |
| 06. Sep 2019 13:09:37 CEST | spankonin | Updated APD batch assignment via API. |
| 02. Sep 2019 12:08:49 CEST | markus.moritz | |
| 08. Jul 2019 14:53:53 CEST | spankonin | Updated APD batch assignment via API. |
| 05. Jul 2019 13:45:51 CEST | jreher | Updated APD location via API. |
| 03. Jan 2019 00:13:10 CET | Tobias | Corrected dark current unit. |
| 24. May 2018 14:39:23 CEST | Tobias | Updated location to "Bochum". |
| 18. May 2018 12:01:31 CEST | Tobias | Imported from Hamamatsu datasheet. |