RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2609029861 - Archived data from Fri, 05. July 2019 13:45:44 CEST


This is all the information about APD 2609029861. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2609029861
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum Box 71
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C12
Break-through voltage: 410 V
Voltage for Gain 100 (T=+25°C): 381.9 V
Dark current: 5.9 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 71
Position in Box: 28
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: none     T = -25 °C: none
Voltage for Gain 150: T = +20 °C: none     T = -25 °C: none
Voltage for Gain 200: T = +20 °C: none     T = -25 °C: none
Gain/Voltage slope at M = 100: T = +20 °C: none     T = -25 °C: none
Gain/Voltage slope at M = 150: T = +20 °C: none     T = -25 °C: none
Gain/Voltage slope at M = 200: T = +20 °C: none     T = -25 °C: none
Break-through voltage: T = +20 °C: none     T = -25 °C: none
   
Notes:


Version history

Time Author Change comment
16. Feb 2021 12:25:01 CET jreher Updated box and position via API.
15. Feb 2021 15:37:14 CET jreher Assigned to capsule by script
05. Feb 2021 10:57:27 CET jreher Updated box and position via API.
03. Feb 2021 18:23:00 CET jreher Updated box and position via API.
11. Aug 2020 17:36:59 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:41:19 CET jreher Updated APD location via API.
25. Nov 2019 17:16:32 CET jreher Updated box and position via API.
06. Sep 2019 13:09:00 CEST spankonin Updated APD batch assignment via API.
02. Sep 2019 11:28:43 CEST markus.moritz
08. Jul 2019 14:53:36 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:45:44 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:09 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:01:26 CEST Tobias Imported from Hamamatsu datasheet.