RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2407026440 - Archived data from Mon, 17. August 2020 15:53:27 CEST


This is all the information about APD 2407026440. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2407026440
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: F12
Break-through voltage: 409 V
Voltage for Gain 100 (T=+25°C): 380.8 V
Dark current: 6.9 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 11
Position in Box: 14
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 12. Nov 2018
Sent for analysis after irradiation: 13. Dec 2018
Return for assembly: none
   
Irradiation:  
Date: 07. Dec 2018
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 10. Dec 2018
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 380.533535 V     T = -25 °C: 345.1805563 V
Voltage for Gain 150: T = +20 °C: 388.3516129 V     T = -25 °C: 352.5559676 V
Voltage for Gain 200: T = +20 °C: 392.6717857 V     T = -25 °C: 356.6875921 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.259797552 V-1     T = -25 °C: 4.94179295 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.495013832 V-1     T = -25 °C: 9.834635306 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.03327239 V-1     T = -25 °C: 15.46081218 V-1
Break-through voltage: T = +20 °C: 408.6441004 V     T = -25 °C: 372.3151768 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:49:10 CET jreher Updated box and position via API.
03. Feb 2021 18:15:18 CET jreher Updated box and position via API.
07. Oct 2020 12:12:57 CEST jreher Assigned to capsule by script
17. Aug 2020 15:53:27 CEST jreher Updated box and position via API.
11. Aug 2020 17:21:32 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:31:19 CET jreher Updated APD location via API.
25. Nov 2019 17:11:45 CET jreher Updated box and position via API.
04. Jul 2019 11:07:41 CEST jreher Updated APD location via API.
12. Jun 2019 16:46:13 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:31:08 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 15:02:50 CEST jreher Updated APD batch assignment via API.
03. Jan 2019 00:12:56 CET Tobias Corrected dark current unit.
13. Dec 2018 13:29:14 CET markus.moritz
30. Nov 2018 13:45:01 CET markus.moritz
12. Jan 2018 13:16:53 CET Tobias Imported from Hamamatsu datasheet.