RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2405026265 - Archived data from Tue, 08. December 2020 17:07:09 CET


This is all the information about APD 2405026265. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2405026265
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2520029078/2405026265
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: red
   
Manufacturer information:  
Wafer position: D05
Break-through voltage: 411 V
Voltage for Gain 100 (T=+25°C): 382.9 V
Dark current: 5.6 nA
   
Screening Logistics:  
Available: No
Storage Box: 9
Position in Box: 12
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 12. Nov 2018
Sent for analysis after irradiation: 13. Dec 2018
Return for assembly: none
   
Irradiation:  
Date: 07. Dec 2018
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 10. Dec 2018
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 383.5014255 V     T = -25 °C: 347.8720808 V
Voltage for Gain 150: T = +20 °C: 391.3610326 V     T = -25 °C: 355.2884816 V
Voltage for Gain 200: T = +20 °C: 395.7320722 V     T = -25 °C: 359.4730145 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.264001173 V-1     T = -25 °C: 4.694167896 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.68043655 V-1     T = -25 °C: 9.297945883 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.48971224 V-1     T = -25 °C: 16.1986745 V-1
Break-through voltage: T = +20 °C: 410.5152808 V     T = -25 °C: 374.1774239 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:48:54 CET jreher Updated box and position via API.
03. Feb 2021 18:15:02 CET jreher Updated box and position via API.
08. Dec 2020 17:07:09 CET jreher Assigned to capsule by script
17. Aug 2020 16:01:01 CEST vwerner Updated box and position via API.
11. Aug 2020 17:21:11 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:30:57 CET jreher Updated APD location via API.
25. Nov 2019 17:11:35 CET jreher Updated box and position via API.
04. Jul 2019 11:06:01 CEST jreher Updated APD location via API.
12. Jun 2019 16:46:03 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:30:58 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 15:02:18 CEST jreher Updated APD batch assignment via API.
03. Jan 2019 00:12:51 CET Tobias Corrected dark current unit.
13. Dec 2018 13:55:54 CET markus.moritz
23. Nov 2018 12:05:50 CET markus.moritz
12. Jan 2018 13:16:50 CET Tobias Imported from Hamamatsu datasheet.