Subdetector specification:  
Serial: 2403026137
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E06
Break-through voltage: 410 V
Voltage for Gain 100 (T=+25°C): 381.3 V
Dark current: 5.9 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 7
Position in Box: 21
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 21. Sep 2018
Sent for analysis after irradiation: 02. Oct 2018
Return for assembly: none
   
Irradiation:  
Date: 27. Sep 2018
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: 28. Sep 2018
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 381.5328008 V     T = -25 °C: 346.0722268 V
Voltage for Gain 150: T = +20 °C: 389.3356646 V     T = -25 °C: 353.450931 V
Voltage for Gain 200: T = +20 °C: 393.6803078 V     T = -25 °C: 357.5923716 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.477444834 V-1     T = -25 °C: 4.906634279 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.797468875 V-1     T = -25 °C: 9.151443733 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.36483432 V-1     T = -25 °C: 16.21470295 V-1
Break-through voltage: T = +20 °C: 409.2037858 V     T = -25 °C: 372.9397488 V
   
Notes:


Version history

Time Author Change comment
27. Oct 2021 14:25:15 CEST oafedulidis Assigned to capsule by script
15. Apr 2021 16:57:56 CEST jreher Updated box and position via API.
05. Feb 2021 10:48:32 CET jreher Updated box and position via API.
03. Feb 2021 18:14:47 CET jreher Updated box and position via API.
11. Aug 2020 17:20:53 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:30:36 CET jreher Updated APD location via API.
25. Nov 2019 17:11:26 CET jreher Updated box and position via API.
04. Jul 2019 11:05:46 CEST jreher Updated APD location via API.
12. Jun 2019 16:45:55 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:30:50 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 15:01:12 CEST jreher Updated APD batch assignment via API.
03. Jan 2019 00:12:47 CET Tobias Corrected dark current unit.
02. Oct 2018 13:48:03 CEST markus.moritz
26. Sep 2018 11:59:05 CEST markus.moritz
12. Jan 2018 13:16:48 CET Tobias Imported from Hamamatsu datasheet.