RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2403026119 - Archived data from Tue, 11. August 2020 17:20:49 CEST


This is all the information about APD 2403026119. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2403026119
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2403026119/2320025822
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: blue
   
Manufacturer information:  
Wafer position: G06
Break-through voltage: 409 V
Voltage for Gain 100 (T=+25°C): 382.7 V
Dark current: 4.4 nA
   
Screening Logistics:  
Available: No
Storage Box: 7
Position in Box: 3
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 21. Sep 2018
Sent for analysis after irradiation: 02. Oct 2018
Return for assembly: none
   
Irradiation:  
Date: 27. Sep 2018
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: 28. Sep 2018
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 382.1848882 V     T = -25 °C: 346.9301082 V
Voltage for Gain 150: T = +20 °C: 390.0214087 V     T = -25 °C: 354.3095872 V
Voltage for Gain 200: T = +20 °C: 394.3762776 V     T = -25 °C: 358.4586563 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.588671937 V-1     T = -25 °C: 5.013041475 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.138478637 V-1     T = -25 °C: 9.35202106 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.33126976 V-1     T = -25 °C: 14.8197073 V-1
Break-through voltage: T = +20 °C: 408.9996238 V     T = -25 °C: 373.5090592 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:48:27 CET jreher Updated box and position via API.
03. Feb 2021 18:14:45 CET jreher Updated box and position via API.
11. Aug 2020 17:20:49 CEST jreher Updated APD characteristics via API.
03. Apr 2020 10:06:40 CEST jreher Assigned to capsule by script
20. Feb 2020 13:34:28 CET jreher Updated box and position via API.
27. Nov 2019 19:30:35 CET jreher Updated APD location via API.
25. Nov 2019 17:11:25 CET jreher Updated box and position via API.
04. Jul 2019 11:05:44 CEST jreher Updated APD location via API.
12. Jun 2019 16:45:54 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:30:49 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 15:01:10 CEST jreher Updated APD batch assignment via API.
03. Jan 2019 00:12:47 CET Tobias Corrected dark current unit.
02. Oct 2018 14:45:00 CEST markus.moritz
26. Sep 2018 11:42:00 CEST markus.moritz
12. Jan 2018 13:16:48 CET Tobias Imported from Hamamatsu datasheet.