RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2509028182 - Archived data from Tue, 11. August 2020 17:31:21 CEST


This is all the information about APD 2509028182. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2509028182
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2408026546/2509028182
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: red
   
Manufacturer information:  
Wafer position: G08
Break-through voltage: 408 V
Voltage for Gain 100 (T=+25°C): 380.9 V
Dark current: 13.6 nA
   
Screening Logistics:  
Available: No
Storage Box: 170
Position in Box: 31
EP1 batch: 24
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 16. Jul 2019
Sent for analysis after irradiation: 19. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 16. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 17. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 380.119393 V     T = -25 °C: 345.2385374 V
Voltage for Gain 150: T = +20 °C: 387.9823612 V     T = -25 °C: 352.5618339 V
Voltage for Gain 200: T = +20 °C: 392.3499468 V     T = -25 °C: 356.6860328 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.704899244 V-1     T = -25 °C: 5.048745499 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.360062123 V-1     T = -25 °C: 9.316657286 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.82466406 V-1     T = -25 °C: 14.93115234 V-1
Break-through voltage: T = +20 °C: 407.6926564 V     T = -25 °C: 371.5730163 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:10:14 CET jreher Updated box and position via API.
11. Aug 2020 17:31:21 CEST jreher Updated APD characteristics via API.
25. Jun 2020 11:22:04 CEST jreher Assigned to capsule by script
19. Feb 2020 11:53:41 CET jreher Updated box and position via API.
27. Nov 2019 19:59:00 CET jreher Updated APD location via API.
25. Nov 2019 17:24:18 CET jreher Updated box and position via API.
30. Jul 2019 15:17:15 CEST spankonin Updated APD batch assignment via API.
19. Jul 2019 17:07:59 CEST markus.moritz
05. Jul 2019 13:58:44 CEST jreher Updated APD location via API.
24. Jun 2019 14:01:29 CEST jgrochowski Updated APD batch assignment via API.
03. Jan 2019 00:12:15 CET Tobias Corrected dark current unit.
12. Jan 2018 13:12:03 CET Tobias Imported from Hamamatsu datasheet.