Show Hamamatsu Avalanche Photo Diode 2505027956 - Archived data from Fri, 05. February 2021 11:11:06 CET
This is all the information about APD 2505027956. If it is wrong, edit the data.
| Subdetector specification: | |
|---|---|
| Serial: | 2505027956 |
| Type: | Hamamatsu Avalanche Photo Diode |
| Detector: | unassigned |
| Unit: | unassigned |
| Preamp: | 0 |
| Current location: | Bochum |
| Installation information: | |
| Label: | none |
| Manufacturer information: | |
| Wafer position: | D08 |
| Break-through voltage: | 408 V |
| Voltage for Gain 100 (T=+25°C): | 379.8 V |
| Dark current: | 5.7 nA |
| Screening Logistics: | |
| Available: | Yes |
| Storage Box: | 177 |
| Position in Box: | 13 |
| EP1 batch: | 19 |
| EP1 batch after irradiation: | none |
| Shipment: | |
| Grid number: | none |
| Position in grid: | none |
| Arrival for irradiation: | 02. Jul 2019 |
| Sent for analysis after irradiation: | 09. Jul 2019 |
| Return for assembly: | none |
| Irradiation: | |
| Date: | 04. Jul 2019 |
| Dose used: | 37 Gy |
| Temperature: | 20 °C |
| Position: | none |
| Bias voltage: | none |
| Annealing: | |
| Date: | 05. Jul 2019 |
| Temperature: | 80 °C |
| Duration: | 48 h |
| Measurement results: | |
| Voltage for Gain 100: | T = +20 °C: 380.3952322 V T = -25 °C: 345.0640917 V |
| Voltage for Gain 150: | T = +20 °C: 388.2529274 V T = -25 °C: 352.4044637 V |
| Voltage for Gain 200: | T = +20 °C: 392.6451773 V T = -25 °C: 356.5271429 V |
| Gain/Voltage slope at M = 100: | T = +20 °C: 4.50546351 V-1 T = -25 °C: 4.967854388 V-1 |
| Gain/Voltage slope at M = 150: | T = +20 °C: 8.839121264 V-1 T = -25 °C: 9.300607314 V-1 |
| Gain/Voltage slope at M = 200: | T = +20 °C: 15.62996956 V-1 T = -25 °C: 14.83597394 V-1 |
| Break-through voltage: | T = +20 °C: 407.6294856 V T = -25 °C: 370.9016107 V |
| Notes: |
Version history
| Time | Author | Change comment |
|---|---|---|
| 15. Apr 2021 16:22:07 CEST | oafedulidis | Assigned to capsule by script |
| 16. Feb 2021 12:28:29 CET | jreher | Updated box and position via API. |
| 05. Feb 2021 11:11:06 CET | jreher | Updated box and position via API. |
| 11. Aug 2020 17:29:54 CEST | jreher | Updated APD characteristics via API. |
| 27. Nov 2019 20:00:15 CET | jreher | Updated APD location via API. |
| 25. Nov 2019 17:24:49 CET | jreher | Updated box and position via API. |
| 18. Jul 2019 10:26:47 CEST | spankonin | Updated APD batch assignment via API. |
| 09. Jul 2019 14:59:46 CEST | markus.moritz | |
| 05. Jul 2019 14:50:13 CEST | jreher | Updated APD location via API. |
| 24. Jun 2019 13:18:05 CEST | jgrochowski | Updated APD batch assignment via API. |
| 03. Jan 2019 00:12:09 CET | Tobias | Corrected dark current unit. |
| 12. Jan 2018 13:12:00 CET | Tobias | Imported from Hamamatsu datasheet. |