RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2414027034 - Archived data from Wed, 03. February 2021 18:16:22 CET


This is all the information about APD 2414027034. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2414027034
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C02
Break-through voltage: 414 V
Voltage for Gain 100 (T=+25°C): 384.8 V
Dark current: 5.9 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 19
Position in Box: 5
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 16. Jan 2019
Sent for analysis after irradiation: 25. Jan 2019
Return for assembly: none
   
Irradiation:  
Date: 21. Jan 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 22. Jan 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 385.0714943 V     T = -25 °C: 349.3472082 V
Voltage for Gain 150: T = +20 °C: 392.9604034 V     T = -25 °C: 356.8502218 V
Voltage for Gain 200: T = +20 °C: 397.351538 V     T = -25 °C: 361.0728375 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.44250525 V-1     T = -25 °C: 4.959328969 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.494238243 V-1     T = -25 °C: 8.938483419 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.24728481 V-1     T = -25 °C: 15.24331166 V-1
Break-through voltage: T = +20 °C: 412.9274128 V     T = -25 °C: 376.6575702 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:50:12 CET jreher Updated box and position via API.
03. Feb 2021 18:16:22 CET jreher Updated box and position via API.
11. Aug 2020 17:22:46 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:32:46 CET jreher Updated APD location via API.
25. Nov 2019 17:12:21 CET jreher Updated box and position via API.
04. Jul 2019 11:08:45 CEST jreher Updated APD location via API.
12. Jun 2019 16:46:48 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:31:43 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 15:09:35 CEST jreher Updated APD batch assignment via API.
25. Jan 2019 14:17:04 CET markus.moritz
03. Jan 2019 00:11:42 CET Tobias Corrected dark current unit.
12. Jan 2018 13:11:47 CET Tobias Imported from Hamamatsu datasheet.