RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2414027025 - Archived data from Wed, 03. February 2021 18:16:18 CET


This is all the information about APD 2414027025. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2414027025
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2414027025/2509028172
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: blue
   
Manufacturer information:  
Wafer position: B09
Break-through voltage: 414 V
Voltage for Gain 100 (T=+25°C): 385.8 V
Dark current: 6.2 nA
   
Screening Logistics:  
Available: No
Storage Box: 18
Position in Box: 46
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 16. Jan 2019
Sent for analysis after irradiation: 28. Jan 2019
Return for assembly: none
   
Irradiation:  
Date: 21. Jan 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 22. Jan 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 386.9041488 V     T = -25 °C: 350.8453925 V
Voltage for Gain 150: T = +20 °C: 394.7471069 V     T = -25 °C: 358.3533194 V
Voltage for Gain 200: T = +20 °C: 399.1012733 V     T = -25 °C: 362.5761698 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.49054503 V-1     T = -25 °C: 4.726901519 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.555635072 V-1     T = -25 °C: 9.227640819 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.32943628 V-1     T = -25 °C: 15.74737182 V-1
Break-through voltage: T = +20 °C: 413.4950486 V     T = -25 °C: 377.3297023 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:50:09 CET jreher Updated box and position via API.
03. Feb 2021 18:16:18 CET jreher Updated box and position via API.
07. Oct 2020 12:18:32 CEST jreher Assigned to capsule by script
17. Aug 2020 15:53:58 CEST jreher Updated box and position via API.
11. Aug 2020 17:22:44 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:32:39 CET jreher Updated APD location via API.
25. Nov 2019 17:12:20 CET jreher Updated box and position via API.
04. Jul 2019 11:08:41 CEST jreher Updated APD location via API.
12. Jun 2019 16:46:47 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:31:42 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 15:09:34 CEST jreher Updated APD batch assignment via API.
28. Jan 2019 11:55:44 CET markus.moritz
03. Jan 2019 00:11:42 CET Tobias Corrected dark current unit.
12. Jan 2018 13:11:47 CET Tobias Imported from Hamamatsu datasheet.