RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1906020691 - Archived data from Tue, 11. August 2020 17:35:17 CEST


This is all the information about APD 1906020691. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1906020691
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C13
Break-through voltage: 406 V
Voltage for Gain 100 (T=+25°C): 377.7 V
Dark current: 11.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 193
Position in Box: 5
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 18. Apr 2019
Sent for analysis after irradiation: 29. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 25. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 27. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 377.8517032 V     T = -25 °C: 342.6974172 V
Voltage for Gain 150: T = +20 °C: 385.7463095 V     T = -25 °C: 350.0292303 V
Voltage for Gain 200: T = +20 °C: 390.1455594 V     T = -25 °C: 354.1809113 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.593063557 V-1     T = -25 °C: 5.026697416 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.950808263 V-1     T = -25 °C: 9.303550471 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 13.82683371 V-1     T = -25 °C: 14.55313047 V-1
Break-through voltage: T = +20 °C: 405.4762942 V     T = -25 °C: 369.3214817 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:13:13 CET jreher Updated box and position via API.
11. Aug 2020 17:35:17 CEST jreher Updated APD characteristics via API.
27. Nov 2019 20:03:11 CET jreher Updated APD location via API.
25. Nov 2019 17:26:03 CET jreher Updated box and position via API.
16. Aug 2019 12:20:45 CEST jreher Updated APD location via API.
09. Aug 2019 16:25:34 CEST jreher Updated APD batch assignment via API.
31. Jul 2019 13:31:09 CEST markus.moritz
23. Jun 2016 11:48:54 CEST Tobias Imported from Hamamatsu datasheet.