RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1901020380 - Archived data from Tue, 11. August 2020 18:09:17 CEST


This is all the information about APD 1901020380. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1901020380
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: G07
Break-through voltage: 406 V
Voltage for Gain 100 (T=+25°C): 377.2 V
Dark current: 12.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 251
Position in Box: 50
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 378.1189326 V     T = -25 °C: 342.7608353 V
Voltage for Gain 150: T = +20 °C: 385.9970239 V     T = -25 °C: 350.0967515 V
Voltage for Gain 200: T = +20 °C: 390.3898666 V     T = -25 °C: 354.2343103 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.536629337 V-1     T = -25 °C: 5.031201441 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.917918856 V-1     T = -25 °C: 9.397781629 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.52819887 V-1     T = -25 °C: 14.74065356 V-1
Break-through voltage: T = +20 °C: 405.6745847 V     T = -25 °C: 369.4190183 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:20:40 CET jreher Updated box and position via API.
11. Aug 2020 18:09:17 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:32:51 CEST jreher Updated APD location via API.
21. Jul 2020 14:32:41 CEST jreher Updated box and position via API.
09. Jul 2020 15:02:14 CEST jreher Updated APD batch assignment via API.
23. Jun 2016 11:48:47 CEST Tobias Imported from Hamamatsu datasheet.