RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1802020073 - Archived data from Tue, 11. August 2020 18:10:16 CEST


This is all the information about APD 1802020073. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1802020073
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: A06
Break-through voltage: 422 V
Voltage for Gain 100 (T=+25°C): 395.2 V
Dark current: 7 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 241
Position in Box: 13
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 394.8590543 V     T = -25 °C: 358.2809677 V
Voltage for Gain 150: T = +20 °C: 402.7988583 V     T = -25 °C: 365.9886728 V
Voltage for Gain 200: T = +20 °C: 407.1916329 V     T = -25 °C: 370.3049812 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.618740768 V-1     T = -25 °C: 4.648396673 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.217382574 V-1     T = -25 °C: 9.230022841 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.43983692 V-1     T = -25 °C: 14.44668918 V-1
Break-through voltage: T = +20 °C: 419.7774769 V     T = -25 °C: 385.020183 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:19:21 CET jreher Updated box and position via API.
11. Aug 2020 18:10:16 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:31:19 CEST jreher Updated APD location via API.
21. Jul 2020 14:31:09 CEST jreher Updated box and position via API.
29. Jun 2020 16:14:25 CEST jreher Updated APD batch assignment via API.
23. Jun 2016 11:47:39 CEST Tobias Imported from Hamamatsu datasheet.