RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1708019022 - Archived data from Tue, 11. August 2020 17:52:46 CEST


This is all the information about APD 1708019022. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1708019022
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D08
Break-through voltage: 418 V
Voltage for Gain 100 (T=+25°C): 389.1 V
Dark current: 4.9 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 281
Position in Box: 28
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 389.8281266 V     T = -25 °C: 353.5167274 V
Voltage for Gain 150: T = +20 °C: 397.7845583 V     T = -25 °C: 361.1738895 V
Voltage for Gain 200: T = +20 °C: 402.1944249 V     T = -25 °C: 365.4730522 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.571932528 V-1     T = -25 °C: 4.670085233 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.068939421 V-1     T = -25 °C: 8.827045028 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.06237876 V-1     T = -25 °C: 16.03982424 V-1
Break-through voltage: T = +20 °C: 417.8937391 V     T = -25 °C: 381.4529718 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:24:36 CET jreher Updated box and position via API.
11. Aug 2020 17:52:46 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:38:55 CEST jreher Updated APD location via API.
21. Jul 2020 14:38:45 CEST jreher Updated box and position via API.
09. Dec 2019 18:29:00 CET spankonin Updated box and position via API.
09. Dec 2019 18:26:25 CET spankonin Updated APD batch assignment via API.
23. Jun 2016 11:42:25 CEST Tobias Imported from Hamamatsu datasheet.